Configure Menu
MENU ITEM
SOFTWARE
Monitor/
NAME
Controller
Label for Anteroom
AnteRm Label
RPM20
Measurements
Display Meas
Displayed
RPM10 and
RPM20
Display Average
Display Avg
RPM10 and
RPM20
Display Units
Units
RPM10 and
RPM20
ITEM DESCRIPTION
The AnteRm Label item allows the user to set the room
number or other designator for the anteroom.
NOTE: AnteRm Label is only active if the # of Rooms
item is set to 1 Room with Anteroom or
2 Rooms with Anteroom.
The Display Meas item selects which measurements will
be presented on the display during normal operating
mode. Use the Units item to choose the units of measure:
ROOM STATUS displays the room mode as negative,
positive or no isolation.
ROOM PRESSURE displays the room mode and the
current measurement of room pressure differential.
ALL displays the room mode and all currently connected
measurements. Only functions when # of Rooms is set to
1 Room
NOTE: Measurements will still enable alarms if not on the
display. The measurement will not appear on the
display
The Display Avg item selects the display's running
average period. The display-averaging period is the length
of time the face velocity has been averaged before being
displayed. The Display Avg item value may be set
between 0.5 and 40 seconds. The higher the averaging
value, the more stable the display.
The Units item selects the unit of measure that the
monitor displays all values (except calibration span).
These units display for all menu items setpoints, alarms,
flows, etc.
ITEM RANGE
DEFAULT VALUE
13 characters of text
ANTEROOM
Room Status,
Room Status
Room Pressure, All
1, 2, 3, 5, 10, 20, or
20 seconds
40 seconds
in H
O, cfm, F
in H
2
Pa, l/s, C
Pa, cmh, C
O, cfm
2