Anterm Alarm Menu - Pressura RPM10 Operation And Service Manual

Room pressure monitor
Hide thumbs Also See for RPM10:
Table of Contents

Advertisement

AnteRm Alarm Menu

SOFTWARE
MENU ITEM
NAME
Anteroom Negative
Neg Hi Alm
High Alarm
RPM20
Anteroom Positive
Pos Low Alm
Low Alarm
RPM20
Anteroom Positive
Pos Hi Alm
High Alarm
RPM20
ITEM DESCRIPTION
The Neg Hi Alm item sets the negative high pressure
alarm setpoint. A high alarm condition is defined as when
the room is more negative than the Neg Hi Alm setpoint.
This item is active when the TSI key switch is in negative
room pressure position or when NEGATIVE is selected in
ROOM MODE item. However, it is always accessible
through the menu system.
The Pos Low Alm item sets the positive low pressure
alarm setpoint. A low alarm condition is defined as when
the room is less positive than the Pos Low Alm setpoint.
This item is active when the TSI key switch is in positive
room pressure position or when POSITIVE is selected in
ROOM MODE item. However, it is always accessible
through the menu system.
The Pos Hi Alm item sets the positive high pressure
alarm setpoint. A high alarm condition is defined as when
the magnitude of the room pressure rises above the
Pos Hi Alm setpoint.
This item is active when the TSI key switch is in positive
room pressure position or when POSITIVE is selected in
ROOM MODE item. However, it is always accessible
through the menu system.
ITEM RANGE
DEFAULT VALUE
-0.19500 in H
O to
-0.10000 in H
2
+0.19500 in H
O
2
Note: Neg Hi Alm
cannot be set
less negative
than the Neg
Lo Alm
-0.19500 in H
O to
0.01000 in H
2
+0.19500 in H
O
2
Note: Pos Low Alm
cannot be set
more positive
than the Pos Hi
Alm
-0.19500 in H
O to
0.10000 in H
2
+0.19500 in H
O
2
Note: Pos Hi Alm
cannot be set
less positive
than the Pos
Lo Alm
O
2
O
2
O
2