Interface Menu
SOFTWARE
MENU ITEM
NAME
Network Address
Nurse
for Nurse's Station
Address
RPM10 and
RPM20
LON Configuration
LON
RPM20
ITEM DESCRIPTION
The Nurse Address item sets the main network address of
the room pressure monitor when communicating with the
Nurse's Station Monitor. Each unit on the network must
have its own unique address.
NOTE: PresSura Model RPM10 and RPM20 monitors will
have rooms displayed on the Nurse's Station
Monitor in order of the Nurse Address. The
PresSura monitor with the lowest Nurse Address
will be displayed at the top-left of the Nurse's
Station Monitor screen. If a PresSura monitor is
configured for more than 1 room, then the rooms
will be displayed on the Nurse's Station in order of
Room 1, Room 2, and Anteroom.
When the SERVICE PIN option is selected, the Model
RPM20 sends a broadcast message containing its Neuron
ID and program ID. This is required to install the Model
®
RPM20 on the LonWorks
network, or to reinstall the
Model RPM20 after using the GO UNCONFIGURED
command.
Selecting the GO UNCONFIGURED option resets the
Model RPM20 monitor's authentication key. This is
required in the event a foreign network tool inadvertently
acquires a Model RPM20 and installs it with network
management authentication. The Model RPM20 monitor's
owner will then be unable to reclaim the Model RPM20
over the network.
NOTE: The LON item is only functional when
Comm Type is set to LON.
DEFAULT
ITEM RANGE
VALUE
1 to 8
1
Service Pin
N/A
Go Unconfigured