Pressura RPM10 Operation And Service Manual page 27

Room pressure monitor
Hide thumbs Also See for RPM10:
Table of Contents

Advertisement

Configure Menu
MENU ITEM
SOFTWARE
Monitor/
NAME
Controller
Enable Access
Passcode
Codes
RPM10 and
RPM20
Rm1 Alarm Menu
SOFTWARE
MENU ITEM
NAME
Mode of Room 1
Room Mode
RPM10 and
RPM20
Room 1 Alarm
Alarm Enable
Enable
RPM10 and
RPM20
ITEM DESCRIPTION
The Passcode item selects whether an access code
(pass code) is required to enter the menu items. The
Passcode item prevents unauthorized access to a menu.
If the Passcode item is:
OFF
no code is required to enter the room
mode or menu screens.
ROOM MODE
access code is required to enter the
room mode screens but not the menu
screens
MENUS
access code is required to enter the
menu screens but not the room mode
screens
ALL
access code is required to enter the
room mode and menu screens.
ITEM DESCRIPTION
The Room Mode item selects the room pressure
direction. This item enables all related alarms, for
pressure direction selected.
NOTE: No Isolation Room Mode can be selected from the
main running screen.
The Alarm Enable item enables the low and high alarm
functions. When this item is entered, the monitor will show
buttons for Low Alarms and High Alarms. Press the button
to toggle between enabling and disabling the alarms.
NOTE: The Alarm Enable item enables or disables
pressure, flow, temperature and humidity alarms.
ITEM RANGE
DEFAULT VALUE
Off
Menus
Room Mode
Menus
All
ITEM RANGE
DEFAULT VALUE
Positive
Negative
Negative
Enabled
Low Alarms
Disabled
Enabled
High Alarms
Disabled

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rpm20

Table of Contents