Aktakom ABM-4082 User Manual page 98

5 1/2 digit multimeter
Table of Contents

Advertisement

ABM-4082
AC Current (True RMS, AC Coupling)
Resolution and Full Scale Reading
Rate
Slow
Med
Fast
(1)
Typical voltage across the input terminals at full scale reading.
Accuracy
Rate
Slow
Med
Fast
(1)
Specified accuracy at input >5% of the range
Measurement method: True RMS measurement
Maximum Crest factor: 3.0 at full scale
Maximum Input and Overload Current Protection: 2A/ 250V fuse.
98
Range
Resolution
0.1μA
10.0000mA
10μA
1.00000A
100μA
10.0000A
0.1μA
10.0000mA
10μA
1.00000A
100μA
10.0000A
1μA
10.000mA
10μA
1.0000A
10.000A
1mA
Range
10~20Hz
10.0000mA
1.0+0.080
1.00000A
1.0+0.080
10.0000A
1.0+0.080
10.0000mA
1.0+0.150
1.00000A
1.0+0.150
10.0000A
1.0+0.150
10.000mA
1.0000A
10.000A
Full Scale
Burden Voltage
Reading
& Shunt Resistor
<0.15V / 10.1Ω
11.9999
<0.3V / 0.1Ω
1.19999
<0.15V / 10mΩ
11.9999
<0.15V / 10.1Ω
11.9999
<0.3V / 0.1Ω
1.19999
<0.15V / 10mΩ
11.9999
<0.15V / 10.1Ω
11.999
<0.3V / 0.1Ω
1.1999
<0.15V / 10mΩ
11.999
(1)
Accuracy (1 year)
20~50Hz
50~2kHz
0.50+0.080
0.25+0.080
0.50+0.080
0.25+0.080
0.50+0.080
0.25+0.080
0.50+0.150
0.25+0.150
0.50+0.150
0.25+0.150
0.50+0.150
0.25+0.150
1.0+0.20
0.5+0.10
1.0+0.20
0.5+0.10
1.0+0.20
0.5+0.10
(1)
2~10kHz
2.0+0.080
2.0+0.080
2.0+0.080
2.0+0.150
2.0+0.150
2.0+0.150
3.0+0.10
3.0+0.10
3.0+0.10

Advertisement

Table of Contents
loading

Table of Contents