Aktakom AM-1016 User Manual

Network multimeter

Advertisement

Quick Links

Network Multimeter
AM-1016
User's Manual
www.tmatlantic.com

Advertisement

Table of Contents
loading

Summary of Contents for Aktakom AM-1016

  • Page 1 Network Multimeter AM-1016 User’s Manual www.tmatlantic.com...
  • Page 2: Safety Information

    1. SAFETY INFORMATION WARNING To ensure safe operation, and in order to exploit to the full functionality of the meter, please follow the directions in this section carefully. This multimeter has been designed according to IEC-1010 concerning electronic measuring instruments with an over voltage category CAT II 1000V, CAT III 600V and pollution2.
  • Page 3: Maintenance

    1.2.3 When the value scale to be measured is unknown beforehand, set the range selector at the highest position. 1.2.4 Before rotating the range selector to change functions, disconnect test leads from the circuit under test. 1.2.5 When carrying out measurements on TV or switching power circuits, always remember that there may be high amplitude voltages pulses at test points which can damage the meter.
  • Page 4: Measuring Voltage

    2. DESCRIPTION The 3 in 1 digital multi-tester has been designed to combine the functions of Digital Multimeter, Telephone Line Tester and Network Cable Tester. • DC voltage measurement, 5 ranges from 200mV to 1000V • AC voltage measurement, 5 ranges from 200mV to 700V •...
  • Page 5: Measuring Current

    3.1.2 Set the rotary switch at the desired "V' range position and connect test leads across the source or load under measurement. 3.1.3 Put down the "AC/DC" when measuring the voltage. Meter will be transformed between DC and AC range. 3.1.4 If you need data hold when measuring, you can put on "HOLD", it will hold the reading;...
  • Page 6 3.4.1 Connect the black test lead to the COM jack and the red test lead to the V/ Ω. jack. (NOTE: The polarity of red lead connection is positive"+") 3.4.2 Set the rotary switch at desired Ω range position and connect test leads across the resistance under measurement.
  • Page 7 The Network Cable Tester will check a fault condition in the above descending order before detecting other fault conditions. The detection and indication of the presence of a fault is handled on a. "one-per- test" basis. Once a fault is corrected, it is recommended the cable be tested again for other faults. OPEN There is no "OPEN"...
  • Page 8: Specifications

    3.7.1 Connect the RJ45 test jack to one end of the cable to be tested. 3.7.2 Connect the remote unit to the other end of the cable. 3.7.3 Push TEST button to perform test. Example: The Cable Fault is a SHORT on Pair 1-2 and Pair 3-6, the LCD status will be as follows: Pair 1-2, Pair 3-6, Pair 4-5, Pair 7-8, SHIE and SHO will display at the same time.
  • Page 9 AC VOLTAGE Range Resolution Accuracy 200mV 0.1mV + 1.2 % of rdg + 3digits 10mV ± 0.8 % of rdg ±3 digits 200V 100mV 700V + 1.2 % of rdg + 5 digits Input Impedance: 10MΩ Frequency Range: 40Hz to 400Hz (for AC) Response: Average, calibrated in rms of sine wave 4.3 CURRENT DC CURRENT...
  • Page 10: Battery And Fuse Replacement

    Overload Protection: F 200Ma/250V fuse for 200μA to 200mA ranges Frequency Range: 40Hz to 400Hz (for AC) Response: Average, Calibrated in rms of sine wave 4.4 RESISTANCE Range Resolution Accuracy 200Ω 0.1Ω ± 0.8 % of rdg ± 3 digits 2KΩ...
  • Page 11 Fuse rarely need replacement and blow almost always as a result of the operator's error. Open the case as mentioned above, and then take the PCB out from the front cover. Replace the blown fuse with same ratings. WARNING Before attempting to open the case, be sure that test leads have been disconnected from measurement circuits to avoid electric shock hazard.