ABM-4082
compared to the "seed" reading during the HOLD process.
:STATe <b>
Command syntax:
[:SENSe[1]]:HOLD:STATe <b>
Command Parameter:
<b> =
0 or OFF
1 or ON
Query:
:STATe?
Description:
This command is used to enable or disable HOLD.
Speed Commands
: NPLCycles <n>
Command Systac:
[:SENSe[1]]:CURRent:AC:NPLCycles <n>
[:SENSe[1]]:CURRent[:DC]:NPLCycles <n>
[:SENSe[1]]:VOLTage:AC:NPLCycles <n>
[:SENSe[1]]:VOLTage[:DC]:NPLCycles <n>
[:SENSe[1]]:RESistance:NPLCycles <n>
[:SENSe[1]]:FRESistance:NPLCycles <n>
Command Parameter:
<n> =
0.1 to 10
DEFault
MINimum
MAXimum
Query:
:NPLCycles?
Description:
The integration period (measurement speed) for the basic measurement functions (except Frequency
and Period) is set using the: NPLCycles command. NPLC (Number of Power Line Cycles) expresses
the integration period by basing it on the power line frequency. For example, for a PLC of 1, the
integration period in seconds would be1/60 (for 60Hz line power) which is 16.67 msec.
:RANGe Commands
:[UPPer] <n>
Command syntax:
76
Disable HOLD
Enable HOLD
Query state of HOLD
Set power line cycles per integration
1
0.1
10
Query programmed NPLC value
Set NPLC for ACI
Set NPLC for DCI
Set NPLC for ACV
Set NPLC for DCV
Set NPLC for Ω2
Set NPLC for Ω4