Sense Subsystem Command - Aktakom ABM-4082 User Manual

5 1/2 digit multimeter
Table of Contents

Advertisement

ABM-4082
6.3.4

SENSe subsystem command

This SENSe subsystem is used to configure and control Model DMM measurement functions and are
summarized in Table 6-4.
Command
[:SENSe[1]]
:FUNCtion<name>
:FUNCtion?
:DATA?
:HOLD
:WINDow <NRf>
:WINDow?
:COUNt <NRf>
:COUNt?
:STATe <NRf>
:STATe?
:CURRent:AC
:NPLCycles <n>
:NPLCycles?
:RANGe
[:UPPer] <n>
[:UPPer]?
:AUTO <b>
:AUTO?
:REFerence <n>
:STATe <b>
:STATe?
:ACQuire
:REFerence?
:AVERage
:TCONtrol <name>
:TCONtrol?
:COUNt <n>
:COUNt?
:STATe <b>
:STATe?
70
Table 6-4
SENSe Command Summary
Function Description
Select
measurement
'VOLTage:DC',
'CURRent:AC',
'PERiod', 'DIODe', 'CONTinuity'.
Query function.
Output the lastest reading
Path to control Hold feature:
Set Hold window(%); 0.01 to 10
Query Hold window
Set Hold count; 2 to 100
Query HOLD count
Enable or disable HOLD
Query state of HOLD
Path to configure AC current
Set A/D integration rate (Line cycle; 0.1 to 10)
Query line cycle integration rate
Path to configure measurement ranges
Select range (0 to 10)
Query range
Enable or disable auto range
Query auto range
Specify reference (-10 to 10)
Enable or disable reference
Query state of reference
Use input signal as reference
Query reference value
Path to configure and control the filter
Select filter type(MOVing or REPeat)
Query filter type
Specify filter count (1 to 100)
Query filter count
Enable or disable filter
Query state of digital filter
function:
'VOLTage:AC',
'RESistance',
'FRESistance',
'CURRent:DC',
'FREQuency',
Default
VOLT:DC
1
5
OFF
1
10
ON
0
OFF
5
OFF