Reading Hold - Aktakom ABM-4082 User Manual

5 1/2 digit multimeter
Table of Contents

Advertisement

ABM-4082
be displayed.
9. Press
to move down a level to set the type of delay mode.
10. Using
or
11. Press Auto (ENTER) to confirm the choice. The message "CHANGE SAVED" will be displayed to show
that the change is now in effect. If AUTO mode is chosen, DMM will exit the parameter level and move
up a level to the command level.
12. If MANU mode is chosen, DMM requires you to continue to specify the delay time. The present delay
time will be displayed.
DELAY: 0000mS
13. Use
and
or decrement the digits. Enter a value of delay time (from 0 to 6000ms).
14. Press Auto (ENTER) to confirm the choice. The message "CHANGE SAVED" will be displayed to
show that the change is now in effect. DMM automatically exits the parameter level and moves up a
level to the command level.
15. Press Shift →
Note:
Changing the delay to MANUAL on one function changes the delays on all functions to
MANUAL.
Measurement Sample
The primary device action is a measurement. However, the device action block could include the following
additional actions:
Filtering — If the repeating filter is enabled, the instrument samples the specified number of reading
conversions to yield a single filtered reading. Only one reading conversion is performed if the filter is
disabled, or after the specified number of reading conversions for a moving average filter is reached. The
output of filter feeds hold function.
Hold — Withhold enabled, the first processed reading becomes the "Seed" reading and operation loops
back within the measurement sample block. After the next reading is processed, it is checked to see if it is
within the selected window (0.01%, 0.1%, 1%, 10%) of the "seed" reading. If the reading is within the
window, operation again loops back within the measurement sample block. This looping continues until the
specified number (2-100) consecutive readings are within the window. If one of the readings is not within the
window, the instrument acquires a new "seed" reading and the hold process continues.
4.2.2

Reading Hold

When a hold reading is acquired as described in "Measurement Sample", an audible beep is sounded (if
enabled) and the reading is considered a "true measurement". The reading is held on the display until an
"out of window" reading occurs to restart the hold process. The reading hold feature allows you to capture
and hold a stable reading on the display.
Take steps below to enable and set reading hold function.
1.
Press Shift + Trig to enable reading hold function
Press Shift →
2.
3.
Use
or
38
to select AUTO or MANU delay mode.
keys to choose a numerical place and use
key to exit from the menu and return to the measurement status.
to enter the menu on the menu level, "A: MEAS MENU" will be displayed.
key to move across to the TRIG MENU on the menu level, "C: TRIG MENU" will
and
keys to increment

Advertisement

Table of Contents
loading