Query:
:UPPer?
:LOWer?
Description:
This command is used to specify the upper and lower limit for LIMIT1. The actual limit depends on
which measurement function is currently selected. For example, a limit value of 1 is 1V for the volts
functions (DCV or ACV), 1A for the current function (DCI or ACI), 1Ω on the ohms functions (2 or 4). A
limit value is not range sensitive. A limit of 1 for DCV is 1V on all measurement ranges.
:STATe <b>
Command syntax:
:CALCulate3:LIMit[1]:STATe <b>
Command Parameter:
<b> =
0 or OFF
1 or ON
Query:
:STATe?
Query state (on or off) of limit test
Description:
This command is used to enable or disable the LIMIT test. When enabled, the limit test will be
performed.
:FAIL?
Command syntax:
:CALCulate3:LIMit[1]:FAIL?
Description:
This command is used to read the results of the LIMIT1 test
0 = Limit test failed
1 = Limit test passed
The response message (0 or 1) only tells you if a limit test has passed or failed. It does not tell you
which limit (upper or lower) has failed.
Query upper limit
Query lower limit
Disable limit test
Enable limit test
ABM-4082
69