Command parameter :
<function> = VOLTage[:DC]
VOLTage:AC
CURRent[:DC]
CURRent:AC
RESistance
FRESistance
FREQuency
PERiod
DIODe
CONTinuity
Query:
:CONFigure?
Description:
This command configures the instrument for subsequent measurements on the specified function.
Basically, this command places the instrument in a "one-short" measurement mode. You then use
the :READ? command to trigger a measurement and acquire a reading.
When this command is sent, the DMM will be configured as follows:
The function specified by this command is selected.
All controls related to the selected function are defaulted to the *RST values.
Continuous initiation is disabled.
The control source of the Trigger Model is set to Immediate.
The count values of Trigger Model are set to one.
The delay of the Trigger Model is set zero.
All math calculations are disabled.
Buffer operation is disabled. A storage operation currently in process will be aborted.
Autozero is set to the *RST default value.
This command is automatically adderted when the :MEASure? Command is sent.
FETCh? Command
Command syntax:
:FETCh?
Description:
This query command is used to obtain the lastest post-processed reading. This command does not
affect the configuration of the instrument.
This command does not trigger a measurement. The command simply requests the last available
reading. This command continues to return the same old reading, until there is a new reading.
DC voltage
AC voltage
DC current
AC current
2-wire resistor
4-wire resistor
Frequency
Period
Diode testing
Continuity test
Query the selected function.