:ENABle <b>
Command syntax:
:DISPlay:ENABle <b>
Command Parameter:
<b> =
0 or OFF
1 or ON
Query:
:ENABle?
Description:
This command is used to enable or disable front panel display circuitry. When disabled, the instrument
operates at a higher speed. While disabled, the display is frozen. All front panel controls except LOCAL
are disabled. Normal display operation can be resumed by using :ENABle command or pressing
LOCAL key to enable the display.
6.3.3
CALCulate Subsystem
The commands in this subsystem are used to configure and control the calculate subsystems and are
summarized in Table 6-3.
Disable front panel dislay
Enable front panel display
Query state of the display
ABM-4082
63