Aktakom ABM-4082 User Manual page 67

5 1/2 digit multimeter
Table of Contents

Advertisement

buffer.
:TRACe Command
:CLEar
Command syntax:
:CALCulate2:TRACe:CLEar
Description:
This action command is used to clear the readings in the buffer. If the buffer is not cleared, a subsequent
store will overwrite the old readings. If the subsequent store is aborted before the buffer becomes full, you
could end up with some "old" readings still in the buffer.
:PIONts <NRf>
Command syntax:
:CALCulate2:TRACe:POINts <NRf>
Command Parameter:
<NRf> =
2 to 512
Query:
:POINts?
Query buffer size.
Description:
This command is used to specify the buffer size for readings saving.
:DATA?
Command syntax:
:CALCulate2:TRACe:DATA?
Description:
When this command is sent and DMM is addressed to talk. All the readings stored in the buffer are sent
to the computer.
:FORMat <name>
Command syntax:
:CALCulate2:FORMat <name>
Command Parameter:
<name> =
NONE
MEAN
SEDViation Standard deviation of readings in buffer
MAXimum
MINimum
Query:
:FORMat?
Specify buffer size.
No calculation
Mean value of readings in buffer
Largest reading in buffer
Lowest reading in buffer
Query programmed math format.