ABM-4082
this command while in any other function causes an error. Also, if the latest reading is overflowed or a
reading has not been triggered, an error occurs when this command is sent.
:AVERage Commands
This command is used to configure and control the filter.
:STATe <b>
Command syntax:
[:SENSe[1]]:CURRent:AC:AVERage:STATe <b>
[:SENSe[1]]:CURRent[:DC]:AVERage:STATe <b>
[:SENSe[1]]:VOLTage:AC:AVERage:STATe <b>
[:SENSe[1]]:VOLTage[:DC]:AVERage:STATe <b>
[:SENSe[1]]:RESistance:AVERage:STATe <b>
[:SENSe[1]]:FRESistance:AVERage:STATe <b>
Command parameter:
<b> =
1 or ON
0 or OFF
Query:
:STATe?
Description:
These commands are used to enable or disable the digital filter for the specified function. When
enabled, readings will be filtered according to how the filter is configured.
:TCONtrol <name>
Command syntax:
:CURRent:AC:AVERage:TCONtrol <name>
:CURRent[:DC]:AVERage:TCONtrol <name>
:VOLTage:AC:AVERage:TCONtrol <name>
:VOLTage[:DC]:AVERage:TCONtrol <name>
:RESistance:AVERage:TCONtrol <name>
:FRESistance:AVERage:TCONtrol <name>
Command parameter:
<name> =
REPeat
MOVing
Query:
:TCONtrol?
Descripton:
80
Enable the digital filter
Disable the digital filter
Query state of digital filter
Select repeating filter
Select moving filter
Query filter type
Control filter for ACI
Control filter for DCI
Control filter for ACV
Control filter for DCV
Control filter for Ω2
Control filter for Ω4
Select filter type for ACI
Select filter type for DCI
Select filter type for ACV
Select filter type for DCV
Select filter type for Ω2
Select filter type for Ω4