These commands are used to select the type of averaging filter for the specified function.
:COUNt <n>
Command syntax:
[:SENSe[1]]:CURRent:AC:AVERage:COUNt <n>
[:SENSe[1]]:CURRent[:DC]:AVERage:COUNt <n>
[:SENSe[1]]:VOLTage:AC:AVERage:COUNt <n>
[:SENSe[1]]:VOLTage[:DC]:AVERage:COUNt <n>
[:SENSe[1]]:RESistance:AVERage:COUNt <n>
[:SENSe[1]]:FRESistance:AVERage:COUNt <n>
Command parameter:
<n> =
1 to 100
DEFault
MIMimum
MAXimum
Query:
:COUNt?
Description:
These commands are used to specify the filter count. In general, the filter count is the number of
readings that are acquired and stored in the filter buffer for the averaging calculation. The larger the
filter count, the more filtering that is performed.
:THReshold Commands
Use these commands to set the maximum range input (signal level) for frequency and period
measurements.
:RANGe <n>
Command syntax:
[:SENSe[1]]:FREQuency:THReshold:VOLTage:RANGe <n>
[:SENSe[1]]:PERiod:THReshold:VOLTage:RANGe <n>
Command parameter:
<n> =
0 to 1010
Query:
:RANGe?
Description:
These commands are used to specify the expected input level. The instrument will then automatically
Specify filter count
10
1
100
Query filter count
Specify signal levels in volts (voltage threshold)
Query maximum signal level.
Specify filter count for ACI
Specify filter count for DCI
Specify filter count for ACV
Specify filter count for DCV
Specify filter count for Ω2
Specify filter count for Ω4
Set voltage threshold range.
Set voltage threshold range.
ABM-4082
81