Advertisement

Quick Links

Advertisement

Table of Contents
loading

Summary of Contents for Aktakom AMM-1009

  • Page 2: Safety Warnings

    SAFETY WARNINGS The following safety information must be observed to insure maximum personal safety during the operation at this meter: Do not use the meter if the meter or test leads look damaged, or if you suspect that the meter is not operating properly.
  • Page 3 Symbols used on this instrument are: Caution: refer accompanying notes. This symbol indicates that the operator must refer explanation Operating Instructions to avoid personal injury or damage to the meter. Caution: risk of electric shock This WARNING symbol indicates a potentially hazardous situation,...
  • Page 4: Symbols And Annunciators

    be connected to a circuit point at which the voltage with respect to earth ground exceeds (in this case) 1000 VAC or VDC. Equipment protected throughout by Double Insulation (Class II) Equipment complies with current EU directives. SYMBOLS AND ANNUNCIATORS Low Battery HOLD Data Hold...
  • Page 5: Specifications

    SPECIFICATIONS Display: 3 3/4 digit liquid crystal display LCD with a maximum reading of 3999. Polarity: Automatic, positive implied, negative polarity indication. Overrange: (OL) or (-OL) is displayed Zero: Automatic Low battery indication: A battery “ BAT”symbol is displayed when the battery voltage drops below the operating level.
  • Page 6 Dimensions: 182 (H) x 82 (W) x55 (D) mm Weight: 375g. Accessories: test leads, 9V battery, gift box with carrying case. DC Voltage Range Resolution Accuracy 400.0m 0.1mV +0.5% of rdg + 1 digits 4.000V 40.00V 10mV 400.0V 100mV 1000V +0.5% of rdg + 1 digits Input Impedance: 10MŸ.
  • Page 7 40.00V 10mV 400.0V 100mV 750V +1.5%of rdg + 4 digits Input Impedance: 10MŸ. Overload protection: 500V dc or ac rms on 400mV range, 1000V dc or 750V ac on all other ranges. DC Current Range Resolution Accuracy 40.00mA 10uA +1.0% of rdg + 1 digits 400.0mA 100uA 10mA...
  • Page 8 Resistance [Ÿ] Range Resolution Accuracy 400.0Ÿ 0.1Ÿ +0.8% of rdg + 4 digits 4.000kŸ 1Ÿ +0.8% of rdg + 2 40.00kŸ 10Ÿ digits 400.0kŸ 100Ÿ 4.000MŸ 1kŸ 40.00MŸ 10kŸ +3.0% of rdg +4digits 400.00MŸ ±[(10% of rdg 100k: -20dgts) +10digits] Open circuit voltage: 0.6V dc(3.0V dc on 400Ÿ...
  • Page 9 Capacitance Cx Range Resolution Accuracy 4.000nF +5.0% of rdg+10 dgts 40.00nF 10pF +5.0% of rdg+ 10dgts 400.0nF 0.1nF +5%of rdg + 10 dgts 4.000uF 40.00uF 10nF Test Frequency: about 400Hz Frequency (Auto-ranging) Ranges: 4kHz, 40kHz, 400kHz,4000kHz Accuracy: + (0.1%rdg + 1dgt) Sensitivity: 200mV (10Hz-4kHz);...
  • Page 10: Operation

    OPERATION Before taking any measurements, read the safety information section. Always examine the instrument for damage, contamination( excessive dirt, grease, etc) and defects. Examine the test leads for cracked or frayed insulation. If any abnormal conditions exist do not attempt to make any measurement. VOLTAGE MEASUREMENT 1.Connect the black test lead to COM jack and the red test lead to V: jack.
  • Page 11: Resistance And Continuity Measurement

    2. For current measurements less than 400mA , connect the red test leads to the mA jack and the black test lead to COM jack. 3. For current measurements over 400mA or greater, connect the red test leads to the 10A jack and the black test lead to the COM jack.
  • Page 12 3. Connect the black test lead to COM jack and the red test lead to V: jack. 4. Touch the probes to the test points. In ohms, the value indicated in the display is the measured value of resistance. In continuity test, the beeper sounds continuously, if the resistance is less than 100:.
  • Page 13: Diode Test

    DIODE TEST WARNING: To avoid electric shock, do not test any diode that has voltage on it. 1. Connect the black test lead to COM jack and the red test lead to V: jack. 2. Set the function switch to the position.
  • Page 14: Frequency Measurement

    FREQUENCY MEASUREMENT WARNING: To avoid electric shock, discharge the capacitor under test before measuring. 1. Set the Function/Range switch to the “Hz” position. 2. Connect the red test lead to the n V jack. and the black test lead to the COM jack. 3.
  • Page 15 2. Connect the red test lead to the V jack and the black test lead to the COM jack. 3. Connect the red test lead to the test point and the black lead to the common buss of the logic circuit.
  • Page 16: Maintenance

    MAINTENANCE WARNING: Remove test leads before changing battery or fuse or performing any servicing. BATTERY and FUSE replacement WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery door. 1. When the batteries become exhausted or drop below the operating voltage, the battery warning symbol will appear in the LCD display.
  • Page 17: Battery Installation

    BATTERY INSTALLATION WARNING: To avoid electric shock, disconnect the test leads from any source of voltage before removing the battery cover. Do not operate the instrument with the battery cover removed 1. Disconnect the test leads from the meter. 2. Open the battery cover by loosening the screw using a Phillips head screwdriver.
  • Page 18 1. Disconnect the test leads from the meter and any item under test. 2. Open the fuse door by loosening the screw on the door using a Phillips head screwdriver. 3. Remove the old fuse from its holder by gently pulling it out.

Table of Contents