User Manual
Start - Class 1/2
Determine appropriate test
configuration
Perform 8.2.1 Test
"Harmonic combination"
Flatcurve and Overswing
Any functional
anomalies?
Yes
Test Completed
FAIL
Record Results
CSW Series
Yes
Class 2
required?
No
Perform 8.2.2 Test
"Frequency Sweep"
Any functional
anomalies?
Perform 8.2.3 Test
"Individual Harmonics/
Interharmonics"
Yes
Any functional
anomalies?
Figure 8-28: IEC 61000-4-13 Test Flowchart Class 1 and 2
Yes
Meister
curve required?
No
No
Yes
Yes
No
California Instruments
Yes
Perform 8.2.4 Test
"Meister Curve"
Any functional
anomalies?
No
Test Completed
PASS
Record Results
155