User Manual
MODE. Program CC. Using the PROGRAM screen set the output current limit value to 8
amps. The system should go into current limit at the program value. Return the current
value to 16.0 amps and press the output on/off button to turn the output off. Disconnect the
load.
8. Repeat steps 4 through 7 but set the output for the following: output voltage = 312 volts,
output range = 230 volts, current limit = 8.0 amps. For step 7, the current limit value can be
set to 4 amps.
9. Repeat steps 4 through 8 for each of the three phases.
In the unlikely event the power source does not pass the functional test, refer to the calibration
procedure in Section 6 or call the Ametek customer service department for further assistance.
38
California Instruments