Table 8-12: Abnormal Frequency Transient - Ametek CSW5550 User Manual

Csw series ac power source
Table of Contents

Advertisement

User Manual
VOLT MIN
This test will set the voltage and frequency to levels defined by Table 8-11 for 5 minutes. The test
will be repeated for Group1 only as indicated in Table 8-11. All Groups will repeat the test using
Table 8-11 for the voltage setting and Table 8-10 for the frequency setting. The  key
(backspace) will terminate the test at any time.
VOLT UNDER
This test will drop the output voltage from 115 volts to 60 volts for 7 seconds.
VOLT SURGE
This test requires 180 volts output. If the power source is set at the low voltage range, the high
voltage range will be selected before the test starts. At the end of the test the AC source will be
switched back to the low range.
The output voltage will surge to 180 volts for 100 ms. followed by drop to 148 volts for 1 sec
before it returns to 115 volts. The  key (backspace) will terminate the test at any time.
FREQUENCY TRANSIENTS (Group 1 only)
This test will set the voltage at 115V and will remain at this voltage throughout the test. The
test will cycle the frequency three times as shown in Table 8-12. Each repeat will start from
sequence 2.
134
Seq. No.
Frequency
1
400
2
480
3
400
4
320
5
400

Table 8-12: Abnormal Frequency Transient

California Instruments
Time
5 minutes
5 sec.
10 sec.
5 sec.
10 sec.
CSW Series