Figure 3-16: Harmonics/Trace Analysis Screen - Ametek CSW5550 User Manual

Csw series ac power source
Table of Contents

Advertisement

User Manual
HARMONICS/TRACE ANALYSIS Screen
The fourth measurement screen is dedicated to the advanced measurements available on the
CSW Series only. This screen is not available on the i Series. The Harmonics/Trace Analysis
measurement screen is a true menu screen offering several user accessible fields. These fields
are used to select the desired acquisition trigger and display mode. The actual data is displayed
whenever the ENTER key is pressed while the cursor is on the VIEW or START field. The
following fields are available on this menu:
Entry
FUNCTION
VIEW
DATA MODE
CSW Series

Figure 3-16: HARMONICS/TRACE ANALYSIS screen

Description
Selects Voltage, Current or Both parameters for display.
Available display modes are TABLE, BAR and TRACE.
TABLE mode:
BAR mode:
TRACE mode:
Selects absolute or relative harmonics display for TABLE and
BAR view modes. In relative mode, all harmonics are shown in
a percentage of the fundamental which is normalized at 100 %.
In absolute mode, the harmonic amplitudes are shown in
absolute volts or amperes.
California Instruments
displays the first 50 harmonics in a
tabular text format.
displays the first 50 harmonics in a
graphical bar chart display.
displays the selected Function in a
time domain (waveform) graphical
display.