User Manual
8.1.1.5 AC TESTS
Note: Prior to test selection the standard and the group selection are required. Use the shuttle to
select the standard and the group if applicable.
8.1.1.5.1 Normal state test
Scroll to the NORMAL STATE AC entry using the up and down cursor keys. Press the ENTER
key to select the NORMAL STATE screens. The screen will appear as shown in Figure 8-3.
The DO160 NORMAL 1 and NORMAL 2 screens have the following tests:
1
2
3
4
5
6
7
8
9
The above tests can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test. For some of these tests, numeric data entry
may be required to define the test number or the modulation rate.
VOLT FREQ MIN
Standard/Group
Voltage
Frequency
Standard/Group
Voltage
Frequency
This test will set the voltage and frequency to levels defined by Table 8-1. The test will last for 30
minutes. The test will be repeated for the Group 2 and 3 using the Voltage setting from Table 8-2
and the frequency from Table 9-1. The key (backspace) will terminate the test at any time.
CSW Series
Figure 8-3: Normal state
VOLT FREQ MIN
VOLT FREQ MAX
VOLT UNBALANCE
WAVEFORM DISTORTION
VOLT MODULATION
FREQ MODULATION
POWER INTERRUPT
VOLTAGE SURGE
FREQ TRANSIENT (group 1)
FREQ VARIATION (group 2 & 3)
RTCA
1
100
3
101.5
380
Table 8-1: Normal Voltage and Frequency minimum
RTCA
1
122
3
120.5
420
Table 8-2: Normal Voltage and Frequency Maximum
Group1
Group2
104
104
105.5
105.5
390
360
Group1
Group2
122
122
120.5
120.5
410
650
California Instruments
Group3
104
105.5
360
Group3
122
120.5
800
125