Ametek CSW5550 User Manual page 176

Csw series ac power source
Table of Contents

Advertisement

User Manual
Number
Message String
-168
"Block data not
allowed"
-200
"Execution error"
-201
"Invalid while in local"
-203
"Command protected"
-210
"Trigger error"
-211
"Trigger ignored"
-213
"Init ignored"
-220
"Parameter error"
-221
"Setting conflict"
-222
"Data out of range"
-223
"Too much data"
-224
"Illegal parameter
value"
-226
"Lists not same length"
-241
"Hardware missing"
-254
"Media full"
-255
"Directory full"
-256
"File name not found"
-257
"File name error"
-283
"Illegal variable name"
176
Cause
Remedy
Block data was sent.
Check programming manual for
correct command syntax
Command could not
Command may be inconsistent with
be executed
mode of operation such as
programming frequency when in DC
mode.
Command issued but
Put instrument in remote state before
unit is not in remote
issuing GPIB commands.
state
Command is locked
Some commands are supported by
out
the unit but are locked out for
protection of settings and are not user
accessible.
Problem with trigger
Unit could not generate trigger for
system.
transient execution or measurement.
Trigger request has
Trigger setup incorrect or unit was not
been ignored.
armed when trigger was received.
Check transient system or
measurement trigger system settings.
Initialization request
Unit was told to go to armed state but
has been ignored
was unable to do so. Could be
caused by incorrect transient system
or measurement acquisition setup.
Parameter not
Incorrect parameter or parameter
allowed.
value. Check programming manual
for allowable parameters
Requested setting
Check other settings. E.g. trying to
conflicts with other
program a DC offset while in AC
setting in effect.
mode
Parameter data
Check programming manual for
outside of allowable
allowable parameter values
range.
More data received
Check programming manual for
than expected
number of parameters or data block
size
Parameter value is
Check programming manual for
not supported
correct parameters
One or more
All lists must be of same length or
transient lists
transient cannot be compiled and
programmed has
executed.
different length.
N/A
N/A
No storage space left
Delete other settings or data to make
to save settings or
room.
data.
Too many waveform
Delete one or more waveforms from
directory entries
waveform memory to make room.
Waveform requested
Check waveform directory for
not in directory
waveform names present.
Incorrect filename
Too many or non ASCII characters
used in waveform file definition.
Variable name illegal.
Use ASCII characters only
California Instruments