Table 8-19: Dips And Interruptions Tests Performed During Run All - Ametek CSW5550 User Manual

Csw series ac power source
Table of Contents

Advertisement

User Manual
RUN ALL
The RUN ALL selection will cause the following automated test sequence suggested by the
standard to be run:
Step
Output in
% of U
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
The user can change the NOMINAL Ut voltage for this. The RUN ALL Command line will change
to ABORT during the test. Selecting ABORT and pressing the ENTER key will terminate the test
at any time and the output voltage will return to the nominal value.
RUN SINGLE
RUN SINGLE command will run a single test once. The Dip or Interrupt test is defined by the
DIP TO, NO CYCLES, and START ANGLE parameters. These parameters must be set before
starting the test. The following is a description of these parameters.
DIP TO:
NO CYCLES:
START ANGLE:
Note: After each individual run, a 10 second delay is inserted.
CSW Series
No of Cycles
T
0
0.5
0
0.5
0
1
0
5
0
10
0
25
0
50
40
0.5
40
0.5
40
1
40
5
40
10
40
25
40
50
70
0.5
70
0.5
70
1
70
5
70
10
70
25
70
50

Table 8-19: Dips and Interruptions Tests Performed During RUN ALL

The dip voltage level as a percentage of the nominal voltage.
The dip duration in cycles.
The start phase angle of the dip
Start angle
Repeat #
(degrees)
0
180
0,45,90
3 at diff ø
45,90,135
3 at diff ø
90,135,180
3 at diff ø
180,225,270
3 at diff ø
270,315,0
3 at diff ø
0
180
0,45,90
3 at diff ø
45,90,135
3 at diff ø
90,135,180
3 at diff ø
180,225,270
3 at diff ø
270,315,0
3 at diff ø
0
180
0,45,90
3 at diff ø
45,90,135
3 at diff ø
90,135,180
3 at diff ø
180,225,270
3 at diff ø
270,315,0
3 at diff ø
California Instruments
Delay
times
between
repeats (s)
3
10
3
10
10
10
10
10
10
3
10
3
10
10
10
10
10
10
3
10
3
10
10
10
10
10
10
143