User Manual
Scroll to the MIL-STD-704 entry using the up and down cursor keys. Press the ENTER key to
select the MIL704 main menu. One of the screens will appear as shown in Figure 8-33. The
voltage mode and setting will define which menu to select. Refer to Section 8.4.1.2
Note: The user has to turn on the Output relay before starting a test and set the steady state
setup for the test.
8.4.1.6 Revision Selection
The default Revision is E. Revisions supported are D and E. The Revision can be changed from
the front panel. Scroll to the REVISION entry using the up and down cursor keys (see Figure
8-33). Use the shuttle to change the selection.
8.4.1.7 AC MODE TESTS
8.4.1.7.1 Steady StateTest
Scroll to the STEADY STATE AC entry using the up and down cursor keys. Press the ENTER
key to select the STEADY STATE screens. The screen will appear as shown in Figure 8-34.
The MIL704 Steady state 1 and Steady state 2 screens have the following tests:
1
2
3
4
5
6
7
The above tests can be selected by scrolling to the highlighted selection using the up and down
cursor keys and the ENTER key to start the selected test.
VOLTAGE
CSW Series
Figure 8-32: Application Menu
Figure 8-33: MIL704 Menu
Figure 8-34: Steady State Menu
VOLTAGE
FREQUENCY
VOLT UNBALANCE
PHASE DIFFERENCE
VOLT MODULATION
FREQ MODULATION
DISTORTION
California Instruments
161