Internal Calibration/Adjustment for CAH Models...

"Internal Calibration" must be set in the Setup menu: setup / wp-1/ param.1 /
CAL./ADJ. / Cal.Int.
The scale housing has a built-in motorized calibration weight.
Example:
Internal calibration and then adjustment (weighing parameters: factory settings
for CAH... models)
1.) Zero the scale.
2.) Start calibration
(e. g., when adjustment
prompt flashes WP).
cal.Int. is displayed for
two seconds.
y The internal calibration
weight is loaded
automatically.
The difference between
the measured value and
the true weight of the
sample will be displayed
with plus/minus signs.
A printout will be
generated if the process
is cancelled using the
( key.
J
--------------------
24.11.2010
10:13
Type
CAH1E-34ED-H
Ser.no.
27163010
Vers.
C2 101.081210
BVers.
01-62-01
--------------------
Internal calibration
Diff. +
0.2 g
Internal
adjustment
Diff. +
0.0 g
--------------------
24.11.2010
10:15
Name:
--------------------
3.) Activate calibration/
adjustment (press the (
key to cancel).
The adjustment weight is
displayed once adjustment
is finished.
A GMP-compliant
printout is generated.