Neutral Measurement Z Nm (Combics 2) - Sartorius Combics CAW1P Operating Instructions Manual

Combics series
Table of Contents

Advertisement

Operation
56
Operating Instructions Combics Indicators
Neutral Measurement
With this application you can measure the length, surface and volume of parts that
have roughly the same specific weight. The o symbol is displayed as the unit (menu
APPLIC.1).
Features
Save the reference weight "wRef" from the weighing platform:
Enter the reference weight "wRef" through the keypad
Enter the factor for calculation "nRef" using the keypad
Enter reference sample weight using a barcode scanner
Measuring with two weighing platforms
Activate Info mode with the I key
Toggle the display between measurement and weight via the w key
The level of accuracy (display resolution) can be set when the calculated
reference weight is applied
Automatic taring of container weight
Setting: Applic. / aut.tare, menu item 3.7
Automatic initialization when the scale is switched on. The indicator is
initialized with the most recently used values for reference sample quantity
"nRef" and reference sample weight "wRef."
Setting: Applic. / aut.strt, menu item 3.8
Exit application, delete parameters
The value of the reference sample weight in the reference memory remains active
until it is deleted via the c key or overwritten or until the application is changed.
The reference sample weight also remains saved after the scale is turned off.
You can assign different functions to the c key for deleting applications. When
you clear applications, you can delete either the data stored for all applications or
just selected data stored for the active application.
Setting: Applic./CLER.CF menu item 3.24
Tare function:
If you store a tare (weight value) by pressing the ) key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting: Applic. / TARE.FNC menu item 3.25.1 (factory setting)
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting: Applic. / TARE.FNC menu item 3.25.2
Restore factory default settings: Applic./Reset menu item 9.1.
Reference weight
In order to calculate the length, surface or volume of a given sample, the average
weight of a reference quantity of the sample must be known (in the example below,
the reference is 1 meter of electrical cable). There are different ways to enter the
reference weight:
Calculating the reference weight
Place the reference quantity (defined by the calculation factor) on the connected
weighing platform and calculate the reference sample weight by pressing the
O key.
or
Place any amount of the sample material on the connected weighing platform,
enter the calculation factor through the keypad, and press the r key to
calculated reference sample weight.
(Combics 2)
Z nM