Sartorius Combics CAW1P Operating Instructions Manual page 105

Combics series
Table of Contents

Advertisement

Example: Output weight value of +1255.7 g
Position 1
2
3
4
+
*
*
*
Position 1:
Plus +, or minus - or space
Position 2:
Space
Positions 3-10: Weight value with decimal point; leading zeros are output as
spaces.
Position 11:
Space
Positions 12-14: Characters for unit of measure, space or ! sign as a symbol
Position 15:
Carriage return
Position 16:
Line feed
Data Output Format with 22 Characters
Normal Operation
1
2 3 4
5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
K
K K K K K + A A A A A A A A A
K
K K K K K
-
A A A A A A A A A
*
*
*
*
*
*
*
K: ID code character, right-justified with spaces
+-: Plus or minus sign
*:
Space
A: Digit or letter (max. 7 characters plus decimal point)
E: Unit symbol (1 to 3 letters followed by 2-0 spaces)
CR: Carriage return
LF: Line feed
Special Codes
1
2 3 4
5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
S
t
a
t
*
*
*
S
t
a
t
*
*
*
S
t
a
t
*
*
*
S
t
a
t
*
*
*
S
t
a
t
*
*
*
S
t
a
t
*
*
*
*:
Space
– –: Final readout
H: Overload
HH: Overload in checkweighing
L: Underweight
L L: Underweight in checkweighing
C: Calibration/Adjustment
5
6
7
8
9
10 11 12 13 14 15 16
1
2
5
5
.
7
*
*
*
*
*
*
*
*
*
*
*
*
*
*
-
-
*
*
*
*
*
*
H
*
*
*
*
*
*
*
H H
*
*
*
*
*
*
L
*
*
*
*
*
*
*
L
L
*
*
*
*
*
*
C
*
*
Operating Instructions Combics Indicators
Data Interfaces
g
*
*
CR LF
*
E E E CR LF
*
E E E CR LF
*
*
*
*
* CR LF
*
*
*
*
* CR LF
*
*
*
*
* CR LF
*
*
*
*
* CR LF
*
*
*
*
* CR LF
*
*
*
*
* CR LF
*
*
*
*
* CR LF
105

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Combics caw2pCombics caw2sCombics cah1Combics caw1s