Clearing the Preload

Deleting the preload when weighing in legal metrology is only possible when the
menu access switch is open.
The "Delete Preload" function (menu item 1.9.9) must be allocated to the J
key (see page 137).
Once the preload has been deleted, the J key must be reallocated back to its
3
original function in the Setup menu, e. g. external calibration/adjustment with
default weights (Setup menu item 1.9).
Procedure
For scales used in legal metrology: Open the menu access switch.
t
Remove the preload weight from the weighing platform.
t
)
hold
Start the "Delete Preload" function.
t
y After a short period of time the preload will be deleted and the indicator will
automatically switch back to weighing mode.
Re-close the menu access switch.
t