Scale Configuration; Service Mode - Sartorius Combics CAW1P Operating Instructions Manual

Combics series
Table of Contents

Advertisement

Scale Confi guration

Service Mode

Purpose
The Service mode enables access to additional menu items in the Setup menu
(setup) which are not displayed when the Service mode is not active. The most
important calibration and adjustment work for the indicator and for the connected
weighing platform can be carried out in the Service menu, e. g. ADC configuration.
When the Service mode is active, an "S" is shown in the top right-hand corner of the
display. To deactivate the Service mode, restart the indicator (turn the indicator off
and back on again).
In Service mode, the Setup menu is expanded with the following parameters after
entering the user password:
S-DATE for entering the next service date
SER.NO for entering the device serial number
MODEL with the model description
S-SQMIN
ALIB.MEM for deleting the Alibi memory
The Setup menu for WP1 and WP2 can be extended to include the following
setting options:
Param1
CAL./ADJ. Calibration, adjustment
Internal linearization (for WP-2 only)
CAL.EXT External linearization with default weights
CAL.E.USR. External linearization with user-defined weights (entered under 1.18)
SET.PREL. Setting the preload
DEL.PREL. Clearing the preload
HND.XT / CAL./ADJ. Entering the adjustment and linearization weights
LIN. WT.1 Entering the lin. weight 1
LIN. WT.2 Entering the lin. weight 2
LIN. WT.3 Entering the lin. weight 3
LIN. WT.4 Entering the lin. weight 4
ADJ.W/O.W Adjustment without weights (entering the characteristic data
of the load cell(s))
NOM.LOAD Nominal load
RESOLUT Resolution
SENSIT.1 Sensitivity in mV/V for cell 1 (or average value for all load cells)
SENSIT.2 Sensitivity in mV/V for cell 2
SENSIT.3 Sensitivity in mV/V for cell 3
SENSIT.4 Sensitivity in mV/V for cell 4
SAVE Save values for 1. 19
GEOG.DAT Adjustment location (geographical data; or alternatively the
gravitational acceleration at the place of installation)
LATITUD Latitude in degrees
ALTITUD Elevation in meters above sea level
GRAVITY. Gravitational acceleration
SAVE. Save values for 1. 20
ADC Configuration
Applying the serial number of the IS weighing platform (verified weighing
platform at WP2)
Apply the serial number
Inactive (standard WP)
Confi guring Weighing Platforms
Operating Instructions Combics Indicators
1.9.
1.9.5
1.9.6
1.9.7
1.9.8
1.9.9
1.18.
1.18.2
1.18.3
1.18.4
1.18.5
1.19.
1.19.1
1.19.2
1.19.3
1.19.4
1.19.5
1.19.6
1.19.7
1.20.
1 20.1
1 20.2
1 20.3
1 20.4
11
12.1
12.1.1
12.1.2
19

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Combics caw2pCombics caw2sCombics cah1Combics caw1s

Table of Contents