Example:
100% of a sample material should be weighed.
Configuration: The "Weighing in Percent" application is selected, and printout has
been set up.
Place empty container on the scale.
t
)
Tare the scale.
t
This is not required if the automatic tare function is active. The tare weight is
saved automatically when you place the container on the platform.
Add reference material to the container in accordance with reference percentage
t
value (in this example, 85 g).
O
Start the calculation of the reference weight by pressing the O key.
t
y The calculation is based on the active net weight value and the reference
percentage value entered
If the weight is too light, an error code is shown in the main display Inf 29.
If this is the case, set the minimum load to a smaller number of digits.
Add additional material until the reference percentage value has been reached
t
(in this example, 100 g)
Operating Instructions Combics Indicators
Operation
67