Classification W (Combics 2) - Sartorius Combics CAW1P Operating Instructions Manual

Combics series
Table of Contents

Advertisement

Classification
With this application, you can determine whether the weight of a given sample lies
within the limits of a defined weight class (Applic.2 menu).
Features
Classification with 3 or 5 weight classes.
Setting: Applic./Applic.2/CLASS./PARAM.2/QTY. menu item 4.8.
Enter the upper class limits using the keypad or by saving weight values from
a load on the platform
Enter the upper limits of weight classes as absolute values or as a percentage of
deviation from the upper limit of Class 1
Setting: Applic./Applic.2/CLASS./PARAM.2/INPUT menu item 4.9.
Activate Info mode via the I key
Toggle the main display between classification display and weight display by
pressing the w key.
Automatic results printout
Setting: Applic./Applic.2/CLASS./PARAM.2/PRINT menu item 4.10.
Automatic taring of container weight.
Setting: Applic. / aut.tare, menu item 3.7
Automatic initialization when the scale is switched on.
Setting: Applic. / aut.strt, menu item 3.8
Exit application, delete parameters
The initialization values remain active until deleted using the c key or overwritten
or until the application is changed. The class limits also remain saved after the scale
is turned off.
You can assign different functions to the c key for deleting applications.
When you clear applications, you can delete either the data stored for all
applications or just selected data stored for the active application.
Setting: Applic./CLER.CF menu item 3.24
Tare function:
If you store a tare (weight value) by pressing the ) key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting: Applic. / TARE.FNC menu item 3.25.1 (factory setting)
A tare value entered manually overwrites a stored tare value (weight value). If you
enter a tare value manually, a tare value (weight value) stored later overwrites the
manually entered value.
Setting: Applic. / TARE.FNC menu item 3.25.2
Restore factory default settings: Applic./Reset menu item 9.1.
Limits
To use the Classification application, you need to enter the delimiters that separate
one class from another. Limits between the individual weigh classes are required for
the classification. The lower limit of Class 1 is defined by the preset minimum load.
The other classes are configured by defining their upper limits. There are two ways
to enter the delimiters for classes 1 through 3 (or 5):
By saving the weight value indicated: Each upper limit value, with the exception of
the highest class, is entered using the keypad or by saving the weight value of a load
on the weighing platform.
By entering a percentage: The upper value of Class 1 is entered using the keypad or
by saving the value indicated. Upper limits for the other classes are defined by
entering a percentage of deviation from the upper limit of Class 1, using the keypad.
W
(Combics 2)
Operating Instructions Combics Indicators
Operation
77

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Combics caw2pCombics caw2sCombics cah1Combics caw1s

Table of Contents