Confi guring Weighing Platforms
Setup Information
24
Operating Instructions Combics Indicators
External Linearization
–
External linearization when weighing in legal metrology is only possible when
the menu access switch is open.
The "external linearization" function must be allocated to the J key (menu
–
item 1.9.6 or 1.9.7).
Once linearization has been completed, the J key must be reallocated back to its
3
original function in the Setup menu, e. g. external calibration/adjustment with
default weights (Setup menu item 1.9).
Procedure
For scales used in legal metrology: Open the menu access switch.
t
(
Zero the weighing platform.
t
Activate the Service mode (see Page20).
t
J
Start linearization.
t
After approximately 2 seconds you will be prompted to place the first
t
linearization weight on the platform.
Place the required weight on the platform.
t
y After a short time the difference between the measured value and the true
weight of the sample will be displayed.
)
Save the linearization weight (cancel using the ( key).
t
y You will then be prompted to place the second linearization weight on the
platform.
Repeat the procedure for all required linearization weights.
t
y After the last linearization weight has been saved you will be prompted to
remove any load from the weighing pan.
Unload the weighing pan.
t
After a short period of time the zero point will be applied automatically and the
t
indicator will automatically switch back to weighing mode.
Re-close the menu access switch.
t