Checkweighing (Combics 2) - Sartorius Combics CAW1P Operating Instructions Manual

Combics series
Table of Contents

Advertisement

Checkweighing
With this application, you can check whether the sample on the weighing platform
matches a target value or lies within a given tolerance range. Checkweighing also
makes it easy to fill sample materials to a specified target weight.
Selection and settings in the Applic. / Applic.2/ CHECK.WG menu.
Features
Enter the nominal or target weight (set point) and the tolerance range delimiters
either using the keypad or by saving the weight value of a load on the platform.
Enter the tolerance limits as absolute values (Min and Max), as a percentage
deviation from the target or as a relative deviation from the target.
Setting: Applic./Applic.2/CHECK.WG/CHECK.RG menu item 4.5.
The target value can be taken over as a weighed value from a weighing
platform, and the upper and lower tolerance limits are defined as a percentage
deviation from the target value (setting 4.5.2). The following percentages can
be selected as the deviation: 0.1%, 0.2%, 0.5%, 1%, 1.5%, 2%, 3%, 5% or 10%,
selection using the r key.
The target value, lower tolerance limit (minimum) and upper tolerance limit
(maximum) can applied as weighed values from the weighing platform (menu
item 4.5.1).
The target value can be applied as weighed values and via asymmetrical percent
limits (menu item 4.5.3).
The target value can be applied as weighed values and via relative weight limits
(menu item 4.5.4).
Target and tolerance limits checked during input; values must conform to:
upper limit > target > lower limit > 1 digit.
Checkweighing range: either 30% to 170% of the target, or from 10% to infinity.
Results are shown on the main display, as a bar graph and LED as well as sent to
control output ports for further processing.
Toggle the main display between "weight value" and "relation to the target" by
pressing the w key. For the limit value, if the weight in the readout is outside
the tolerance range, "LL" (too low) or "HH" (too high) is displayed
Activate Info mode via the I key
Automatic results printout (Applic./Applic.2/CHECK.WG/CHECK.RG menu
item 4.6).
Automatic taring of container weight (APPLIC./aut.tare menu item 3.7).
Automatic initialization when you switch on the scale with most recently saved
application data (APPLIC./aut.Strt menu item 3.8).
Combics 2 Only:
You can assign different functions to the c key for deleting applications. When
you clear applications, you can delete either the data stored for all applications or
just selected data stored for the active application.
Setting: Applic./CLER.CF menu item 3.24
Tare function:
If you store a tare (weight value) by pressing the ) key, you can later enter a tare
value manually. The tare value you enter is added to the stored tare value.
Setting: Applic. / TARE.FNC menu item 3.25.1 (factory setting)
H
(Combics 2)
Operating Instructions Combics Indicators
Operation
69