Calibration and Adjustment
j
z
Weighing
i
Statistics
o
;
;
Calibrate balance
CAL-Extern
CAL-Intern
Report
Date and Time:
2021-08-10
Calibration function:
Internal calibration
Start calibration:
Start: manually
Calibration deviation:
Dev -0.02 g
Adjustment function:
Internal adjustment
Adjustment deviation:
Dev 0.00 g
m
74
User Manual Secura
®
, Quintix
8.1 Calibration/Adjustment Using an Internal
Calibration Weight (Secura
The balance is fitted with an internal calibration weight in the housing . This
calibration weight is automatically placed on the balance by a motor during internal
calibration and adjustment .
Ensure that the balance is stable, the draft shield is closed (if applicable) and the
t
weighing pan is empty .
Select
A
t
Select the menu key at the bottom left of the display in the current application .
t
y The menu appears .
Mixing
Components
Select the CAL button to access the calibration function .
t
y The
Calibrate balance
CAL-Intern
Select
t
y Calibration begins .
The internal weight is automatically placed on the balance in the housing .
The balance is calibrated and then immediately adjusted .
The internal weight is automatically removed from the balance .
y A report appears on the display .
The report indicates the deviations identified during calibration . The report also
13:56
gives the result of the adjustment process .
Close the report window: Select
t
y The balance is now calibrated and adjusted .
a
Information on potential errors can be found in the "Status Messages" chapter on
page 91 .
®