Sartorius Combics CAW1P Operating Instructions Manual page 101

Combics series
Table of Contents

Advertisement

Connecting a printer
The COM1 standard interface, the optional UniCOM universal interface or both can
be used as a printer interface.
Use as a communications interface
The data protocol can be set to the following operating modes for operation as a
communications interface:
SBI (factory setting)
XBPI-232
XBPI-485 (UniCOM only)
MP8 binary (COM1 only)
SMA
Profibus
Ethernet
Both communication interfaces can be operated independent of each other
(e.g. data transfer and control via a PC using the optional UniCOM interface with
simultaneous printing via the COM1 printer interface). During data communication,
each interface can use a different protocol, e. g. Com1 with SMA and UniCom with
XBPI.
In SBI operation, an indicator and connected weighing platform can be controlled
using ESC commands from the PC via the communications interface (COM1 or
UniCOM) (see page 105).
Each device can only be configured once for two interfaces (COM1 and UniCOM).
Device include:
WP-2
Analog output
If you attempt to reconfigure a device that has already been configured to another
interface (e.g. WP2 to COM1) a second time (to UniCOM), the error message INF 74
appears.
Data Interfaces
Operating Instructions Combics Indicators
101

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Combics caw2pCombics caw2sCombics cah1Combics caw1s

Table of Contents