Example 1:
Checkweighing samples with a target weight of 1250 g and a tolerance range from
-10 g to +30 g. The tolerance values should be entered as absolute values (lower and
upper tolerance limit).
Configuration: The "Checkweighing" application is selected using the setting
INPUT / TAR.MN.MX, a printout has been set up (see "Configuration").
O
t
y The target value symbol flashes at the top of the display.
t
O
t
y The minimum symbol flashes at the top of the display.
1240
t
O
t
y The maximum symbol flashes at the top of the display.
1280
t
O
Save the upper limit value.
t
y Because the sample with the target weight is still on the weighing platform,
t
t
Start target value and tolerance entry using the O key.
Place a sample with the target weight (in this example, 1250 g) on the platform.
Save the target value.
Enter lower limit value (in this example, 1240 g).
Save the lower limit value.
Enter upper limit value (in this example, 1280 g).
the weight is shown on the display with the checkweighing bars. The green LED
indicates a value in the target range.
Remove the sample with the target weight from the platform.
The samples can now be placed on the platform and checked one after the
other.
Operating Instructions Combics Indicators
Operation
73