Competence series; gem plus series; cp series; gc series; gp series; electronic semimicro- and analytical precision balances and precious metal scales (88 pages)
Contents 17 Overview of the Setup Menu in Service Mode 04 Overview Input of the service password Service Concept Setup - Device - WP1 General Information WP1 - Internal - Calibration / Adjustment Different to Combics1/2 Setup - Device - WP1 / COM1 / COM2 Accompanying Literature Setup-Menu for A/D Converter Configuration 06 Description of the Equipment...
Overview General Service Concept Prerequisites for performing maintenance and repair work on Sartorius Combics scales requires considerable experience with both indicators and weighing platforms. In case of defects, repairs are performed on site. Generally, the equipment is not replaced. General Information –...
Mechanical and electrical service or repair work on the complete Combics scale requires considerable experience, and for this reason should be performed only by Sartorius technicians trained at the factory. Any attempt to perform repair work can result in damage to the equipment.
Description of the Equipment The complete Combics scales from Sartorius are primarily used in industry. Because there are a number of hardware configurations to choose from, Combics models are constructed according to customer requirements. The complete Combics scale consists of: –...
The description of the operating Operating Design design is divided into the following sections: – Data Input – Display Modes – Error Codes – Data Output – Saving Data Data Input 65571-010-03.eps There are a number of options for entering data: With Combics 3 you can Before you begin, you need to –...
( – Zero the scale c – Press briefly: Quit application Keypad Input – Cancel a calibration/ program, delete input character Labeled Keys adjustment procedure – Press and hold: Delete entire These keys always have the function input string indicated by the label, but the ) –...
$ Deleting entire input string: Go one level higher Function Keys (Soft Keys) Press and hold c (> 2 sec) The current function of a given soft Show items under selectedentry key is indicated in the last line on the Text Input Through the Keypad display (footer).
$ Entering a space: Input Through the Weighing Platform Input Through the Weighing Platform Input Through the Weighing Platform Input Through the COM Port Input Through the COM Port Input Through the COM Port Input Through the Weighing Platform Input Through the Weighing Platform Input Through the COM Port Input Through the COM Port Press the 0 key...
Activating the xBPI Protocol Activate the xBPI protocol to perform adjustment with the service software Sartocas software or Psion server without activating the BPI mode through the SBI/BPI key. Note: This menu item is accessible in the standard menu. Setup Device parameters COM 1 port Data communications...
Activating the Service Mode Turn on the Combics The weighing instrument is currently in an application mode (such as Weighing or, as in this example, Counting). Enter the service password (see Appendix) and press M to confirm. Note: The number shown in the illustration on the left is not the service password. SuS001_d.bmp The device in now is Service mode.
Open the menu. The Device Device Device Device submenu is opened. Device Select the desired menu item from the next level and open the next submenu. Repeat this procedure until the desired menu item in the lowest menu level can be opened.
Service Menu When the Service mode is activated, the service technician The CAL key function and Activate ext. adj. submenus can access additional menu items that are not visible when of the „Calibration/Adjustment“ menu also contain additi- the Service mode is deactivated. onal menu items.
Page 15
A diagram of the menu tree for the Setup menu in Service mode is shown on the following pages. As mentioned above, which menu items are shown depends on the weighing platform(s) connected. Note: To configure the A/D converter and to enter or change parameters in the „Calibration/Adjustment“...
Setup menu. available strain-gauge load cell or analog Sartorius CAPP, CAPS, IU or IF weighing platform by selecting or entering Parameters not listed above are not affected by your choice parameters in the Setup program.
Overview of the Setup Menu in Service Mode o = Factory setting x = User-defined setting Input of the Setup Application parameters service password Fn key function Device parameters („Device parameters“ menu structure depicted below) Info Language Factory settings: all parameters Setup Device (Device parameters)
Page 18
WP-1 WP-1 WP-1 Internal Internal Internal Calibration/Adjustment Calibration/Adjustment Calibration/Adjustment WP-1 WP-1 Internal Internal Calibration/Adjustment Calibration/Adjustment isoCAL function o Off Adjustment prompt Activate ext. o Activated adjustment3) Deactivated Parameter for Cal./adj. weight external weight Linearization weight 1 Linearization weight 2 Linearization weight 3 Linearization weight 4 Adjust without Input parameters...
Page 19
Setup Setup Setup Device Device Device see previous page see previous page see previous page Setup Setup Device Device see previous page see previous page (Device parameters) Enter the service COM-1 o Off password for access o RS-232 SBI standard SBI verifiable o IS-232 3) (see menu „WP1- RS-232- IS-232“).
Setup-Menu for A/D Converter Configuration Page 1 Access Setup in Service mode WP1 Internal - ADU configuration Standard- Ranges Single range mode Scale interval d configuration Max. cap. Multiple -interval mode Scale interval d Range 1 Range 2 Range 3 Max.
Page 21
Setup-Menu for A/D Converter Configuration Page 2 WP1 - Intern - ADU configuration o Verifiable Accuracy o Class Ranges Singel range mode Verfication scale interval e å / ï class Min.cap Max.cap Multi-interval mode Verfication scale interval e Min. cap Range 1 Range 2 Range 3...
Key to Model Designations Here is an example of how to put together Complete Combics scale with indicator 1 (CW3P...); order numbers. CW3P1-60 FE-LCE means With one load cell (...1...); the following: A maximum, single-range weighing capacity of 60 kg (...60...);...
CW3S4-1500RR-L, our example Complete Combics stainless steel scale of a complete, stainless steel scale, with indicator 2 (CW3S…); means the following: With four load cells (…4…), A maximum, single-range weighing capacity of 1,500kg (…1500…); A platform size of 1,500+1,500 mm (…RR…);...
Page 24
Sartorius AG. The status of the information, specifications and illustrations in this manual is indicated by the date given below. Sartorius AG reserves the right to make changes to the technology, features, specifications, and design of the equipment without notice.
Need help?
Do you have a question about the CW3P and is the answer not in the manual?
Questions and answers