Operation
Available Parameter Settings
( (
Minimum Load for Initialization
52
Operating Instructions Combics Indicators
Entering the reference piece weight
The reference piece weight (i.e., the weight of one piece) can be entered using the
keypad and save with the O key.
Reading the reference piece weight
The reference piece weight can be read using a barcode scanner.
h
The entered value remains active until deleted by pressing the c key or until
overwritten by a new value. It remains saved after the scale is switched off.
Preparation
Open the APPLIC. / APPLIC.1 / COUNT. menu.
t
* = Factory setting
MIN.INIT Minimum load for initialization
RESOLUT Resolution for calculation of reference value
SAVE WT. Parameter for saving weight values
REF.UPDT Reference sample updating .
REF.WP Reference weighing instrument
)
To save the setting, press the ) key.
t
To exit setup: Press the ( key several times.
...
t
You can set the minimum load here, i.e. the load that must be placed on the
weighing platform in order to carry out the application. If the load on the platform
is too light, the following will occur:
Error code Inf 29 appears,
The weighing platform is not initialized,
The preset reference sample quantity is saved.
Setting: Applic./Applic.1/Count./Min.init menu item 3.6.
The minimum load can be set in 10 steps from 1 to 1000 digits (see available
parameters). The "digits" here refer to the scale intervals for the connected weighing
platform. If the interval of the connected platform is 1 g, for example, and 1000 digits
are required, you must place at least 1000 g (= 1000 intervals = 1000 digits) on the
weighing platform for initialization.
1 digit*
1 digit
2 digits
2 digit
5 digits
5 digit
10 digits
10 digT.
20 digits
20 digT.
50 digits
50 digT.
100 digits
100 diG.
200 digits
200 diG.
500 digits
500 diG.
1000 digits
1000 dI.
Display accuracy*
DISP.ACC
Display accuracy + 1 decimal place
10 FOLD
Display accuracy + 2 decimal places
100FOLD
With stability*
STABIL.
With increased stability
ACC.STAB
Off
OFF
Automatic*
automat
No weighing platform selected
NO WP
Weighing platform WP1
wp 1
Weighing platform WP2
wp 2
3.6
3.6.1
3.6.2
3.6.3
3.6.4
3.6.5
3.6.6
3.6.7
3.6.8
3.6.9
3.6.10
3.9.
3.9.1
3.9.2
3.9.3
3.11
3.11.1
3.11.2
3.12.
3.12.1
3.12.3
3.13.
3.13.1*
3.13. 2
3.13. 3