Sartorius Combics CAW1P Operating Instructions Manual page 111

Combics series
Table of Contents

Advertisement

Example with 2 transactions:
HEADER LINE1
HEADER LINE2
14.01.2010
09:43
--------------------
G#
+
1.400 kg
T
+
0.200 kg
N
+
1.200 kg
n
1
G#
+
3.400 kg
T
+
0.200 kg
N
+
3.200 kg
n
2
Single printout (menu setting 3.17.2)
Complete standard printout
configuration is printed for each
transaction.
Example: print 2nd translation
HEADER LINE1
HEADER LINE2
14.01.2010
09:43
--------------------
G#
+
2.400 kg
T
+
0.200 kg
N
+
2.200 kg
n
2
Standard printout
The transaction counter is not printed.
Example: print 2nd translation
G#
+
2.400 kg
T
+
0.200 kg
N
+
2.200 kg
Print menu parameters:
All active sub-items of the currently
displayed menu are printed:
--------------------
MENU
SETUP
WP1
--------------------
1
1.1
1.1.2
1.2.1
1.3.2
...
1.18
1.18.1
CAL.
10.000 kg
etc.
GMP-compliant printouts
Linearization printout
--------------------
14.01.2010
13:00
Type
CAW2P
Ser.no.
12345678
Vers.
C2 100.280810
BVers.
01-62-01
--------------------
Linearization
Wt.1 +
7.00 kg
Wt.2 +
15.00 kg
Wt.3 +
22.00 kg
Wt.4 +
30.00 kg
completed
--------------------
14.01.2010
13:02
Name:
--------------------
Calibration/adjustment printout
--------------------
14.01.2010
13:50
Type
CAW2P
Ser.no.
12345678
Vers.
C2 100.280810
BVers.
01-62-01
--------------------
External calibration
Tar.
+
30.00 kg
Diff. -
0.03 kg
External adjustment
Diff. +
0.00 kg
--------------------
14.01.2010
13:52
Name:
--------------------
Setting the preload printout
--------------------
14.01.2010
13:50
Type
CAW2P
Ser.no.
12345678
Vers.
C2 100.280810
BVers.
01-62-01
--------------------
Setting the preload
completed
--------------------
14.01.2010
13:52
Name:
Data Interfaces
--------------------
Clearing the preload printout
--------------------
14.01.2010
13:50
Type
CAW2P
Ser.no.
12345678
Vers.
C2 100.280810
BVers.
01-62-01
--------------------
Clearing the preload
completed
--------------------
14.01.2010
13:52
Name:
--------------------
Weighing printout with multiple results
(Example with 2 results):
--------------------
14.01.2010
09:43
Type
CAW2P
Ser.no.
12345678
Vers.
C2 100.280810
BVers.
01-62-01
--------------------
HEADER LINE1
HEADER LINE2
14.01.2010
09:43
--------------------
G#
+
2.40 kg
T
+
0.20 kg
N
+
2.20 kg
--------------------
HEADER LINE1
HEADER LINE2
14.01.2010
09:44
--------------------
G#
+
3.40 kg
T
+
0.30 kg
N
+
3.10 kg
--------------------
14.01.2010
09:45
Name:
--------------------
Operating Instructions Combics Indicators
111

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Combics caw2pCombics caw2sCombics cah1Combics caw1s