74
Operating Instructions Combics Indicators
y The LEDs next to the display indicate the results:
yellow LED: sample too heavy
green LED: sample in tolerance range
red LED: sample too light
p
Print the results.
t
Note: If automatic printout of results is enabled, you do not need to press the
p key. The results are printed automatically.
Printout configuration: see page 93
Target
Minimum
Maximum
Gross weight
Tare weight
Net weight
Percentage of deviation from target*
Absolute deviation from target
* When displayed in relation to target value:
If the weight is lighter than the target, the display shows: LL
If the weight is heavier than the target, the display shows: HH
Example 2:
Checkweighing samples with a target weight of 1250 g and a tolerance range
from –10 g to +30 g.
The tolerance values should be entered as a relative deviation from the target value.
Configuration: The "Checkweighing" application is selected using the setting
INPUT / TAR.TOL, a printout has been set up (see "Configuration").
O
Start target value and tolerance entry using the O key.
t
y The target value symbol flashes at the top of the display.
Place a sample with the target weight (in this example, 1250 g) on the platform.
t
O
Save the target value.
t
y The minimum symbol flashes at the top of the display.