Data Interfaces
Analog UniCOM Interface (optional)
Connection Options
100
Operating Instructions Combics Indicators
Level:
Power supply:
Factory setting:
Connection:
Optional UniCOM universal data interface
1)
XBPI operating mode: 9600 baud, 8 data bits, parity: odd, 1 stop bit
2)
Only with the standard COM1 interface
3)
Network address is valid only in XBPI-RS485 operating mode
4)
The following printers can be connected to COM1 and UniCOM
YDP20 (user-defined interface parameters)
YDP14IS (strip or label printer)
YDP04IS (strip or label printer)
Universal printer (user-defined transmission parameters)
The following devices can also be connected to COM1:
Foot / Hand switch
PC (RS-232 interface)
2nd weighing platform (Combics 2 only, RS-232 interface)
External checkweighing display (stop light) via a digital I/O (Sartorius standard)
The following devices can also be connected to the optional UniCOM:
PC (RS-232 interface)
2nd weighing platform (Combics 2 only, in RS-232 or RS-485 mode)
2nd printer (external power source required)
Remote display
Current interface (4-20 mA, 0-20 mA, 0-24 mA, 0-10 V)
Profibus
Ethernet
Galvanically isolated digital I/0
3
You may need to use an external power supply to operate peripheral devices.
Connecting a second weighing platform:
A second weighing platform can be connected to the Combics 2 at COM1, UniCOM
or COM-WP.
COM1 is operated in the RS-232 mode. A second weighing platform can use the
following operating modes:
SBI standard
SBI verifiable
XBPI-232 (factory setting)
ADC-232
UniCOM can be operated in either RS-232 or RS-485 mode. A second weighing
platform can use the following operating modes:
SBI (RS-232 mode)
XBPI-232 (RS-232 mode)
ADC-232 (RS-232 mode)
IS-485 (RS-485 mode, XBPI mode, factory setting)
ADC-485 (RS-485 mode)
4 to 20 mA, 0 to 20 mA, 0 to 24 mA, 0 to 10V
Internal
4 to 20 mA
CAWxP, CAISL2 devices (IP44 protection):
25-pin D-Sub socket
CAWxS, CAH1 devices (IP69K protection):
Connection via screw terminals in the housing,
cable routed into the housing via a cable gland.