Chapter 3
Serial Configuration
Noread Message
Format: <Kkstatus,message>
status:
0 = Disabled
1 = Enabled
Example: To enable Noread Message and enter the message "FAIL," enter
<Kk1,FAIL>.
Bar Code Output
Format: <Klstatus,when to output>
status:
0 = Disabled
1 = Enabled
Example: To set Bar Code Output to End of Read Cycle, enter <Kl1,1>.
Good Decode Reads
Format: <Kmnumber>
number of reads:
Any number from 1 to 31
Example: To change Number of Reads to 3, enter <Km3>.
Match Code
Format: <Kntype,sequential matching,match start position,match length,wild
card character,sequence on noread,sequence on every mismatch>
type:
0 = Disabled
1 = Enabled
2 = Wild Card
3 = Sequential
match length
Default is 1.
1 to 31
3-10
MS-7100/7180 Industrial Scanner User's Manual
message:
Any ASCII string up to 7 digits.
Default is NOREAD.
when to output:
0 = As Soon As Possible
1 = End of Read Cycle
sequential
match start
matching
position
0 = Increment
Default is 0
1 = Decrement
0 to 31
wild card character
Default is *.
Any ASCII character.