Automatic Correction Of Reference Mass - RADWAG WPY User Manual

Hide thumbs Also See for WPY:
Table of Contents

Advertisement

26.2.1. Automatic correction of reference mass

Working mode <
Automatic correction of reference mass>, that can be used for
<
correcting the unit mass <SMP>. To enable the function in parameters you
need to:
• Enter parameter group <
24 of this manual, choose: „
correction of reference mass" and set appropriate option.
Options:
-
No
-
Yes
Function <
Automatic correction of reference mass> in mode
Counting Pieces> is enabled at the moment of estimating the sample
<
quantity and signalled by displaying <PCS> and <SMP> (single piece mass)
on the top part of the display.
There are four criteris of working "Automatic correction of reference
mass" function:
1. equilibrium should be reached,
2. quantity of pieces should be increased,
3. added quantity of pieces should not be greater than double
number of pieces on the pan,
4. the new sample can be different from the old sample by ± 0.3
of pcs (absolute value),
If a user recognises that the sample quantity is adequate the unit mass
(single piece mass) can be saved (see ch. 26.6 of this manual)
and disable the function by pressing ENTER/PRINT.
Notice:
While the function is active ENTER/PRINT changes its functionality.
Pressing ENTER/PRINT does not result in printing and saving weighings.
It concerns the <
Parts counting> comprises a special function
Automatic correction of reference mass disabled
Automatic correction of reference mass enabled
Counting pieces> working mode
Working modes> according to ch.
Counting Pieces /
94
Automatic