RADWAG WPY User Manual page 155

Hide thumbs Also See for WPY:
Table of Contents

Advertisement

33.4.8. Send the stable result in current unit
Syntax: SU CR LF
Possible answers:
SU_A CR LF
SU_E CR LF
SU_I CR LF
SU_A CR LF
MASS FRAME
Frame format:
1
2
3
S
U
space
Example:
S U CR LF – computer command
S U _ A CR LF - command accepted and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command done, mass
value in current unit is returned.
33.4.9. Send the result immediately in current unit
Syntax: SUI CR LF
Possible answers:
SUI_I CR LF
MASS FRAME
Frame format:
1
2
3
S
U
I
Example:
S U I CR LF – computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed and mass
returned
- command accepted and in progress
- timeout while waiting for stable results
- command comprehended but cannot be executed
- command accepted and in progress
- mass value in current unit is returned
4
5
stability
space
- command comprehended but cannot be executed
- mass value in current unit is returned immediately
4
5
stability
space
6
7-15
sign
mass
space
6
7-15
16
sign
mass
space
155
16
17
18
unit
17
18
19
unit
19
20
21
CR
LF
20
21
CR
LF

Advertisement

Table of Contents
loading