RADWAG WPY User Manual page 156

Hide thumbs Also See for WPY:
Table of Contents

Advertisement

33.4.10. Switch on continuous transmission in basic unit
Syntax: C1 CR LF
Possible answers:
C1_I CR LF
C1_A CR LF
MASS FRAME
Frame format:
1
2
3
S
I
space
33.4.11. Switch off continuous transmission in basic unit
Syntax: C0 CR LF
Possible answers answers:
C0_I CR LF
C0_A CR LF
33.4.12. Switch on continuous transmission in current unit
Syntax: CU1 CR LF
Possible answers:
CU1_I CR LF
CU1_A CR LF
MASS FRAME
Frame format:
1
2
3
S
U
I
- command comprehended but cannot be executed
- command comprehended and in progress
- mass value in basic unit is returned
4
5
stability
space
- command comprehended but cannot be executed
- command comprehended and executed
- command comprehended but cannot be executed
- command comprehended and in progress
- mass value in current unit is returned
4
5
stability
space
6
7-15
16
sign
mass
space
6
7-15
16
sign
mass
space
156
17
18
19
20
unit
CR
17
18
19
20
unit
CR
21
LF
21
LF

Advertisement

Table of Contents
loading

Table of Contents