26.4. Setting a reference unit by weighing a sample

Procedure:
• Enter mode <
of this manual,
• If pieces are weighed in a container it needs to be put on the pan
and tarred,
• Press
<Reference Quantity> together with the screen keyboard,
• Enter a value and confirm it by pressing
message is displayed: <Put pieces: xx> (where xx – the value
entered before),
• Put the declared quantity of pieces on the pan and when the result
is stable (symbol
• The program automatically calculates reference unit mass and
causes starting <
Notice:
• The total mass of all pieces put on the weight pan cannot be greater
than the weighing range;
• The total mass of all pieves put on the weight pan cannot be less than
the value declared in parameter „Minimum reference mass" (see
section 26.2.2). If the condition above is not fulfilled the scale dispalays
the message <Too low sample mass>;
• The mass of one unit may not be less than 0.1 of reading division
of the scale. If this condition is not fulfilled the scale dispalays the
message: <Too low piece mass>.
26.5. Setting the reference mass by entering single piece mass directly
to the database
After selecting a product from the product database a mass of single
piece from the field <Mass> is used.
Parts counting> according to ch. 26.1
(estimate piece mass), then the editing field is displayed
) confirm it by pressing
Parts counting >.
, then the following
96
.