Configuration of communication can be configured in:
"
/
Notice:
In submenu <
The detailed description of cooperation scale – barcode scanner can be
found in APPENDIX F in this manual.

16.3.1. Port for barcode scanner

Procedure:
• Enter <
Barcode reader /
Barcode scanners can be connected to:
• RS232 (1),
• RS232 (2).

16.3.2. Prefix / Suffix

Users can edit a prefix <
adjust the program to accept transmission frames from the scanner.
Notice:
A special protocol is required in order the code be received by RADWAG
equipment. It is required to program an appropriate prefix and suffix.
Prefix – one byte 01 hexadecimally, suffix one byte 0D hexadecimally.
The detailed description of cooperation scale – barcode scanner can
be found in APPENDIX F in this manual.
Procedure:
• Enter <
• Chose parameter <
keyboard, a required value (hexadecimal) and confirm it by pressing
.
Devices /
Barcode reader".
Communication> set the baud rate (default 9600b/sec).
Devices> according to ch.16 of the manual, choose
Prefix> or / and suffix <
Barcode Scanner> according to ch.16.3 of the manual,
Prefix> and then enter, using the screen
Port" and then set the appropriate value.
50
Suffix> in order to