No/No; Table 5-7: M9110Ah O - Teledyne 9110AH Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

TELEDYNE Model 9110AH NO
Table 5-7: M9110AH Operating Modes
Mode
ZERO CAL D
ZERO CAL A
ZERO CAL R
ZERO CAL M
SPAN CAL D
SPAN CAL A
SPAN CAL R
SPAN CAL M
M-P CAL
DIAG ELEC
DIAG OPTIC
DIAG OZONE
DIAG AOUT
DIAG
DIAG I/O
DIAG RS232
SETUP x.x
SAMPLE ZS
SAMPLE Z
SAMPLE S
SAMPLE A
SAMPLE

5.4.1 NO/NO

/NO
x
This is the most common mode of operation. It provides continuous output of NO, NO
NO
gas concentrations. To select switching mode press SETUP-MORE-VARS-ENTR and
2
press NEXT to scroll to the MEASURE-MODE variable. Set the value of this variable to NO
NO
, then press ENTR. Press EXIT to return to SAMPLE mode.
x
Analyzer Operator Manual, 01620, Rev. F
X
Description
Automatic dynamic zero calibration
Automatic zero calibration
Remote zero calibration
Manual zero calibration
Automatic dynamic span calibration
Automatic span calibration
Remote span calibration
Manual span calibration
Manual multi-point calibration
Electrical diagnostic test
Optical diagnostic test
Ozone generator diagnostic test
D/A output diagnostic test
Main diagnostic menu
Signal I/O diagnostic
RS232 output diagnostic
Setup mode (x.x is software version)
Sampling; automatic dynamic zero and span calibration enabled
Sampling; automatic dynamic zero calibration enabled
Sampling; automatic dynamic span calibration enabled
Sampling; automatic cal. Enabled
Sampling; automatic cal. Disabled
Switching Mode
2
5-23
and
x