Table 2-1: Final Test Andc - Teledyne 9110AH Instruction Manual

Nitrogen oxides analyzer
Table of Contents

Advertisement

TELEDYNE Model 9110AH NO
Table 2-1: Final Test and Calibration Values
TEST Values
RANGE
NOISE
SAMP FLW
OZONE FL
PMT
AUTOZERO
HVPS
DCPS
RCELL TEMP
BLOCK TEMP
BOX TEMP
PMT TEMP
CONV TEMP
RCEL PRES
SAMP PRES
PMT Volts
NO Conc
NO
Conc
x
PMT Volts
NO Conc
NO
Conc
x
(table continued)
Analyzer Operator Manual, 01620, Rev. F
X
Observed
Units
Value
ppm
ppm
cc/min
cc/min
mV
mV
V
mV
o
C
o
C
o
C
o
C
o
C
IN-Hg-A
IN-Hg-A
Electric Test &Optic Test
Electric Test
mV
ppm
ppm
OPTIC TEST
mV
ppm
ppm
2-13
Nominal Range
5-5000
0.0 - 0.2
300
50 (Default)
550
50 (Optional)
250
15
0-5000
-10 to +50
400 - 700 constant
2500
200
50
2
50
2
8-48
7
1
700
10 (Std)
315
5 (Moly)
2 - 10 constant
25 - 30 constant
2000
200
250
25
250
25
100
20
12.5
2
12.5
2
Reference Section
5.3.4
9.1.1, Table 9-1, 9.2.5
9.3.7, Table 9-1
9.3.6
9.3.8
4.1
9.3.8.5
9.3.5
9.3.8.2
9.3.4.1
9.3.4.1
9.3.8.4
9.3.4.1
9.3.7
9.3.7
9.1.3.2
9.1.3.2
9.1.3.2
9.1.3.3
9.1.3.3
9.1.3.3