Teledyne 9110AH Instruction Manual page 80

Nitrogen oxides analyzer
Table of Contents

Advertisement

TELEDYNE Model 9110AH NO
To print records from a DAS data channel enter:
D REPORT "name" RECORDS=nnn COMPACT|VERBOSE
Examples of reports are:
D REPORT "CONC" RECORDS=35 VERBOSE
D REPORT "CALDAT" RECORDS=10
D REPORT "PNUMTC" RECORDS=155 VERBOSE
Automatic RS-232 reporting can be independently enabled and disabled for each Data Channel.
For all default data channels, automatic reporting is initially set to "OFF." If this property is
turned on, the Data Channel will issue a report with a time and date stamp to the RS-232 port
every time a data point is logged. The report format is shown below:
D 94:08:00 0200 CONC : AVG NXCNC1 = 1234.5 PPB
D 94:08:00 0200 CONC : AVG NOCNC1 = 1234.5 PPB
D 94:08:00 0200 CONC : AVG N2CNC1 = 1234.5 PPB
One CONC report consists of:
D
94:08:00
0200
CONC
AVG
NXCNC1 = 1234.5 PPB
Analyzer Operator Manual, 01620, Rev. F
X
= Type of report (Diagnostic)
= Time and Date stamp (Julian day, Hr, Min)
= Instrument ID number
= Data Channel name
CONC = concentration data
PNUMTC = pneumatic parameters
CALDAT = calibration parameters
= Type of data
AVG = average reading
INST = instantaneous reading
= Name of the parameter
NX = NOx,
NO = NO
N2 = NO2.
5-38